Example.—A thermocouple and leads have a resistance of 5 ohms and are subject to alterations amounting to 1 ohm. To find the errors resulting when indicators of resistances 800, 400, and 50 ohms respectively are used.

From Ohm’s law, C = E/R, the variation in C, with E constant, will be 1 in 805, 1 in 405, and 1 in 55 respectively. As the indications are proportional to the current, the alterations caused will be approximately ⅛ per cent., ¼ per cent., and 2 per cent. The first two may be ignored; the last may be quite serious and lead to the failure of an operation.

It will be seen from the foregoing that low-resistance indicators should only be used for fixed thermocouples and short leads not subject to temperature changes, or, in other words, in a circuit of fixed resistance.

The resistance of an indicator, when unknown, may be found by the following method, suggested by the author:—A resistance box is joined at one end to one terminal of the indicator. To the other terminal a fairly stout iron wire, 18 inches long, is connected, and a similar length of constantan wire is coupled to the other end of the resistance box. The free ends of the wires are twisted into a junction which is dipped into boiling water. The deflection obtained with no resistance in the box (D1) is noted, and resistances (R) are then unplugged until the deflection (D2) is approximately one-half of D1. The resistance (G) of the indicator, ignoring that of the wires, is then given by the formula

    D2R
G   =   ——————
    D1 - D2

as may readily be proved from Ohm’s law, E being constant. This method is extremely simple and reasonably accurate.

Reliable indicators are now procurable from many instrument-makers at a comparatively small cost, progress in this direction having been most marked in recent years, particularly in the case of pivoted instruments. The most convenient form for workshop use is made with an edgewise scale (fig. 14) and may be placed in a suitable position fixed to a bracket. The flat-scale pattern is preferable for use on a laboratory table, or for a portable pyrometer. The sector pattern is also good for workshop use, the dial being visible from a distance.

Fig. 14.—Indicator with Edgewise Scale

Standardizing of Indicators to read Temperatures directly.—The temperature scale of an indicator, for use with a given thermal couple, is always marked by the maker in the case of instruments furnished with a pointer, and, generally speaking, is correct within reasonable limits. It is customary and necessary to send with the instrument a statement of the cold-junction temperature for which the markings are correct; say 20° C. or 60° F. The user should then endeavour to maintain the cold junction at this specified temperature when taking a reading, or otherwise a considerable error may be introduced. It is highly desirable, however, that the user should be able to perform the standardizing himself, if only for checking purposes; and when using a mirror galvanometer as indicator it is necessary to standardize on the spot at which the instrument is fixed. Ability to prepare a temperature scale is further useful, inasmuch as any good millivoltmeter, of range 0 to 20 millivolts, may be used for thermo-electric work of all kinds, and may be calibrated for different junctions, a suitable series resistance being added to enable E.M.F.’s higher than 20 millivolts to be measured. Such an instrument may thus be made extremely useful, both in the workshop and laboratory.

Standardization may be effected either by subjecting the hot junction to several known temperatures, and noticing the deflections corresponding thereto; or by measuring the electromotive force developed by the junction, and calculating the corresponding temperature from a formula which is known to hold for the range comprehended by the instrument. The former method is simpler; and if carefully conducted is quite accurate. The latter method possesses the advantage that readings in millivolts may be translated directly into temperatures when the constants of a given thermal couple are known. It is now usual to mark indicators with a double scale, one reading millivolts and the other temperatures.

Standardization by Fixed Points.—Taking any millivoltmeter which, with a maximum of 20 millivolts at the terminals, will give a full scale deflection, the first step is to arrange that the pointer (or spot of light) shall just remain on the scale at the highest temperature to be attained by the junction. This may be done by placing the hot junction in boiling water and noting the deflection obtained, either in millivolts or equal arbitrary divisions, and also the temperature of the cold junction. The deflection observed is due to a difference of temperature (100-t) deg. C, where t is the temperature of the cold junction. If the highest temperature to be measured is 10 times (100-t), the deflection should be rather less than 110 of the scale, and similarly for any other required temperature limit. If the observed deflection exceed this proportion, a series resistance should be added until the correct value is obtained. This resistance is then permanently installed in the circuit for use with the junction under trial.

Before proceeding further it is necessary to consider whether the pyrometer is to possess a single cold junction of ascertainable temperature (as in fig. 6), or whether it will be arranged with two cold junctions in the head, as in fig. 4. In the former case it is simpler to prepare a “difference” scale; that is, one which reads differences of temperature between the hot and cold junctions, from which the temperature of the hot end may be obtained by adding to the difference that of the cold junction. In the latter case the cold end should be kept by artificial means at the temperature likely to be attained in practice—say 25° C.—a water-bath being suitable for this purpose. It is advisable to remove the shield of the pyrometer when standardizing, so as to expose the hot junction, as closer readings can then be taken.

A number of materials—preferably cheap—of known boiling points or melting points are then selected from a table of fixed points (page 16) so as to give about six points, distributed fairly evenly over the scale. As an example, if it were desired to prepare a temperature scale from 0° to 1000° C., the following might be chosen:—

 
Substance and Condition. Temperature.
Water at boiling point   100° C.   212° F.
Tin at melting point   232   449
Zinc at melting point   419   786
Antimony at melting point   631  1167
Common salt at melting point   800  1472
Copper at melting point (covered with graphite)   1084  1983
 

The hot junction is allowed to attain these temperatures successively, and the corresponding deflection in each case is noted. It is then possible to divide up the whole of the scale to read temperatures directly.

The first reading is taken by placing the junction in a vessel of boiling water, and for a locality near sea level it is not necessary in ordinary work to take account of fluctuations in the boiling point due to alterations of atmospheric pressure. To ensure that the other readings are taken when the substances are exactly at the melting point, the procedure is as follows: about 2-3 lb. of the substance are melted in a salamander crucible, and a small fireclay tube, closed at one end, is inserted in the molten mass. The hot junction is placed in the fireclay tube, and the intervening space filled with asbestos fibre. Great care must be taken not to let the junction touch the fused substance. The crucible is now allowed to cool, and a reading of the deflection taken every half-minute. When the substance is exactly at its solidifying point—identical in general with the melting point—the deflection remains stationary for several consecutive readings, owing to the liberation of latent heat of fusion in sufficient quantity to balance the loss by radiation. This stationary reading is noted for each substance, and represents the deflection given when the hot junction is at the temperature corresponding to the melting point, and the cold junction or junctions at the temperature existing when the observation is made. For melting the materials, a Davies furnace with a large Teclu or Meker burner is convenient up to 850° C.; but to melt the copper a blast lamp is requisite. The molten mass may be allowed to cool in the furnace.

From these observations a calibration curve may be drawn either for differences between hot and cold junctions, or for a steady temperature of the cold junctions. Two sets of data are appended to illustrate the procedure.

 
Temperature
of Hot
Junction.
Pyrometer 1. Iron-constantan. Pyrometer 2.
(Series resistance in Platinum-iridioplatinum.
galvanometer circuit.)  
Deflection. Cold Junction. Difference. Deflection. Cold Junction.
  100° C. 8·9     15° C.    85° C. 5·5    ┐
  232 21·8     17   215 15·6    │ Constant
  419 40·6     19   400 29·4    ├  at
  631 63·8     19   612 45·5    │ 25° C.
  800 83·0     20   780 59·0    │
 1084 . . .      . . .    . . . 82·0    ┘
 

Fig. 15.—Calibration Curves for Two Thermo-electric Pyrometers.

Fig. 15, A, is a calibration curve for thermocouple 1, connecting deflections with corresponding differences between the temperatures of the hot and cold junctions. In order to read from this curve the temperature of the hot end, the reading corresponding to the observed deflection is added to the existing temperature of the cold junction. Thus if a deflection of 56 divisions were obtained with the cold junction at 25°, the temperature of the hot junction would be (540 + 25) = 565° C. The advantage of this method of calibration is that it is unnecessary to take precautions to keep the cold junction at a steady temperature; and when a single cold junction is used, as in fig. 6, this plan should always be followed. It will be noted that this curve passes through zero, as no deflection represents no difference of temperature.

Fig. 15, B, represents the calibration curve for pyrometer 2, and is such that direct readings may be obtained corresponding to any given deflection, for a cold junction temperature of 25°. This curve, therefore, cuts the axis of zero deflection at 25°, as no deflection corresponds to the condition when both hot and cold junctions are at 25°. This method of calibration may be used with advantage for couples of the type shown in fig. 4, where two cold junctions exist in the head, and the simple rule of adding the cold junction temperature does not apply. Many suggestions have been made for correcting for alterations in the temperature of the cold end of such a couple, but none are accurate, and it is necessary to keep this part at the temperature of standardization to secure correct readings. In both of the above calibrations the galvanometer used possessed a scale divided into 100 equal arbitrary divisions.

In making permanent temperature scales from these curves to attach to the existing galvanometer scale, intervals of 100° may be taken and marked opposite to the corresponding divisions on the existing scale. Each 100° may then be equally subdivided into as many parts as the length of scale permits, and numbered at suitable intervals. If the junction used yield a calibration curve departing greatly from a straight line, every 50° interval should be taken, or, if necessary, every 25°. In the examples given both curves are nearly straight lines in the working region, viz. 400° to 800° for the iron-constantan junction, and 500° to 1100° for the platinum-iridioplatinum.

One precaution necessary in standardizing an indicator by this method is to ensure that the metals used are pure, as impurities lower the melting points. If ordered as “pure” from any dealer of repute, the metals will generally be found satisfactory. The common salt used should be the ordinary salt sold in blocks, and not a prepared table salt. A second precaution, when observing melting points, is to guard against a possible error due to the substance becoming “surfused” or “overcooled”; in which case the temperature falls below the ordinary freezing point before solidification commences. When freezing occurs, however, the temperature rises to and remains at the true melting point, and an increase of deflection following a gradual fall always indicates overcooling. The higher deflection then attained is the true freezing point. Antimony frequently overcools to 600° before freezing, but on setting rises to the correct figure—631°. All metals and salts are liable to overcooling occasionally.

Standardization by Measurement of E.M.F.—It has been found, as the result of experiments, that the relation between the E.M.F. developed by a junction and its temperature—under constant conditions of the cold junction—may be expressed approximately by a formula as under:—

log E = A log t + B (Holman’s formula),

where E = electromotive force in microvolts, t = temperature in Centigrade degrees, and A and B are constants depending upon the junction. With certain junctions this formula may be applied over the working part of the scale with an error not exceeding 2° C., but with others the discrepancy is greater. In order to determine the constants A and B, it is necessary to measure the E.M.F. at two known temperatures, which should be chosen as far apart as possible in the working region. When these constants are known, a measurement of E enables the temperature t to be found by calculation.

Example.—Le Chatelier found that a junction at the temperature of melted aluminium (657° C.) gave 6200 microvolts; at the melting point of copper in air (1062° C.) the figure was 10580. Applying in the above formula

   log 6200 = A log 657 + B
and
   log 10580 = A log 1062 + B,

the value of A is 1·2196 and of B 0·302, as may be found by taking logarithms and solving for A and B.

The values of the constants A and B vary for different junctions, and also for different melts of what are reputed to be the same materials. When once determined for a quantity of homogeneous wires, to which the formula applies with sufficient accuracy, it is evident that an indicator with a millivolt scale may be made to read temperatures directly without any necessity for further experiment, although it is always advisable to take one check reading at a fixed point in the working range.

Fig.16.—Potentiometer Method of Measuring E. M. F.

In order to determine the E.M.F. of a junction at different temperatures, the potentiometer method is used, in which the E.M.F. of the test-couple is balanced against the known E.M.F. furnished by a constant cell. The circuit is shown in fig. 16, in which B is an accumulator which sends a current through the resistances R1, R2, and the calibrated wire DE. The cold ends of the couple are attached at P so as to be in opposition to B, and in this branch of the circuit are included a sensitive galvanometer G and a portion of the wire DE. A standard cadmium cell, S, is connected between R1 and R2 at one end, and may be put in circuit with the galvanometer through the switch A. In commencing, S is connected to the galvanometer and R1 adjusted until no deflection is obtained on G. The switch A is now moved over to the circuit of the couple, and the terminal F moved along the wire until zero deflection is again obtained. The E.M.F. of the couple is determined from the relation

E of junction   Resistance of DF
————————   =   ————————
E of standard cell   R2

By exposing the hot end of the junction to successive standard temperatures, and maintaining the cold ends at a known constant temperature, the necessary data for inclusion in a formula may be obtained.

In fixing a permanent temperature scale, calculated from the formula, to a millivoltmeter, it must be remembered that the values given by the experiment are absolute, and independent of the resistance of the circuit composed of the thermo-element and galvanometer. On the other hand, a millivoltmeter is marked to read difference of potential at its terminals; and if in series with a junction and leads of notable resistance, its indications will not be the E.M.F. of the junction. An example will make this point clear.

Example.—A millivoltmeter has a resistance of 100 ohms, and is marked to read P.D. at its terminals. A thermocouple and leads connected to the millivoltmeter have a resistance of 5 ohms. To find the relation between the true E.M.F. of the junction and the readings of the indicator.

If E = the E.M.F. developed by the junction, and V, the reading of the millivoltmeter, = P.D. at its terminals, then the current in the circuit = E105 = V100; and V = (100105)E. That is, the readings are lower by 5 per cent. than the true E.M.F. of the junction. In the same way a low resistance voltmeter, if applied to a cell of high resistance, shows a lower reading than the E.M.F. of the cell.

This example indicates how a table connecting true E.M.F.’s with reading in millivolts may be calculated when the resistances concerned are known. It is presumed, in preparing a scale in this manner, that the resistance of the couple will not be subject to such alterations as to affect the reading.

The advantages of this method of calibration are manifest when a number of junctions are being made from a given batch of wires, as it is only necessary to divide the scale of the indicator so as to represent millivolts—a simple operation—and then to attach a temperature scale. This procedure is much more expeditious than standardizing each indicator at several fixed points when a number are concerned, but for a single junction the fixed point method is easier. The potentiometer method of measuring E.M.F. may also be used to determine temperatures in place of an indicator, and is of great service in cases where very accurate readings are specially required, being far more delicate in detecting small differences of temperature than an indicator. Special potentiometers for thermo-electric work are made by the Cambridge and Paul Instrument Company, Siemens, and others, and are useful in conducting accurate research, but are too elaborate for workshop or ordinary laboratory practice.

Cold Junction Compensators.—The necessity for paying attention to the cold junction has led to various attempts to compensate automatically for changes of temperature at this part of the pyrometer. A thermometer located near the cold junction, as in fig. 6, is all that is needed to correct a two-junction circuit; but when a three-junction circuit is used a correct reading is not secured by adding the excess temperature of the thermometer over the calibration temperature to the reading on the indicator. In Bristol’s arrangement a mercury thermometer, with a large bulb and wide stem, is stationed at the cold junction, and participates in any temperature change. In the stem is placed a loop of thin platinum wire, which forms part of the pyrometer circuit. When the mercury is heated it expands up the stem and short-circuits a portion of the loop, thereby diminishing the resistance of the pyrometer circuit, and tending to increase the deflection on the indicator. Simultaneously the cold junction will be heated, tending to diminish the current, and so to cause a less deflection. By adjustment these two tendencies may be counterbalanced, so that the reading is unaffected, but such adjustment will only apply to a given E.M.F., and therefore to one temperature of the hot junction. Hence this method fails in general application.

Peake’s compensated leads are intended to remedy cold-junction errors by transferring this junction, in effect, to the galvanometer. They are used for pyrometers in which the platinum metals are employed, and consist of wires of two different alloys of copper and nickel, which connect the cold end to the indicator. These alloys are such that the electromotive forces set up at the junctions in the head—Pt and Cu-Ni 1, and Pt-Ir with Cu-Ni 2—are equal and opposite at all working temperatures, and hence changes at the cold junctions do not affect the reading. At the indicator, however, temperature changes would cause an alteration in deflection; but as the indicator is generally placed well away from the furnace, and is not liable to notable heating or cooling, the possible errors are greatly reduced by the use of these leads. They are obviously of no value for use with base-metal pyrometers, as the wires used in such may be prolonged to the indicator, with an identical result.

Fig. 17.—Darling’s Compensator, fitted to Galvanometer.

An automatic compensator for use with base-metal pyrometers has been devised by the author, and is illustrated in figs. 17 and 18. A spiral made of a compound strip of two metals is attached to the needle of the indicator, and coils or uncoils when cooled or heated, thereby moving the pointer over the scale. The length of the spiral is such that an alteration of a given number of degrees in its temperature moves the pointer by the same number of degrees on the scale—or, in other words, the temperature scale of the pyrometer is identical with that of the spiral. The metals forming the junction are continued, in the form of wires, to the interior of the galvanometer, where a cold junction is formed, which will always possess the same temperature as the spiral. The scale is constructed to represent differences of temperature between the hot and cold junctions, and before coupling up the pyrometer the pointer indicates the temperature of the spiral; that is, of the cold junction. On connecting the thermocouple the pointer is moved by the coil of the indicator through an amount represented by the difference in temperature between the two junctions, and therefore finally indicates the temperature of the hot junction.

Fig. 18.—Indicator fitted with Darling’s Compensator.

Example.—If the cold junction were at 20°, the pointer, before connecting the couple, would indicate 20° on the scale. If the hot junction were 580° hotter than the cold, then on completing the circuit the pointer would move 580 additional degrees along the scale, so that the figure indicated would be (20 + 580) = 600°, the temperature of the hot junction. If now the indicator were heated by 10°, the spiral would tend to augment the deflection by 10°, but simultaneously the deflection due to the junctions would fall off by 10°, and the reading would still be 600°.

This method of compensation renders the readings independent of the cold junction, and, in addition to its use for high temperatures, enables ordinary and low temperatures to be read simply and correctly, as will be shown later. The spiral is located in the tower rising from the top of the indicator in fig. 18.

In Paul’s method of compensation the thermocouple and indicator are placed across a Wheatstone bridge, two arms of which contain resistances of copper, whilst the resistances in the other two arms are of manganin. Any change in temperature at the cold junction is shared by these four resistances, and, whilst affecting the resistance of the copper parts, no change is caused in the manganin parts, as this alloy has a negligible temperature coefficient. If, therefore, the bridge were initially balanced at 20° C., and the temperature rose to 30°, the increased resistance of the copper would destroy the balance, and permit of a small current passing through the indicator. A fall to 10°, by diminishing the resistance of the copper, would cause an equal current to pass through the indicator in the opposite direction. The amount of this current is arranged so as to add the rise in temperature of the cold junction to the reading of the indicator in the one case, and to subtract the fall in the other, thus retaining true readings for the cold-junction temperature at which the couple was standardized.

Constant Temperature Cold Junctions.—If the cold junction can be kept at a steady temperature, compensators are unnecessary, but no good practical means of achieving this end has yet been devised. Water-cooled heads have already been referred to; but in many situations the connecting-pipes entailed would be objectionable, and hence this arrangement is not greatly used. An alternative method, suggested by Prof. A. Zeleny, is to bury the cold junction in the ground. Recent experiments, conducted at Cambridge by R. S. Whipple, showed that a junction buried 10 feet deep did not vary in temperature by more than 2° C. over a period of three years. This has led to the adoption of buried junctions in special cases; but it is probable that much greater variations would be experienced in the ground beneath large furnaces, in which case the advantages of this procedure would be lost. A common workshop method is to locate the cold junction in a thermos flask filled with oil, when a temperature constant to 2° C. may be secured, although the changes in the temperature of the surrounding atmosphere may be as great as 150 C. For special work, ice may be used in the thermos flask, thus securing absolute constancy; but this procedure is not feasible in ordinary works practice.

Special-Range Indicators.—When the working range of a pyrometer is from 600° C. upwards, it is evident that the part of the scale occupied by the first 600° is useless, and that it would be an advantage if the whole scale could be utilised for the special working range, so as to secure more exact readings. This may be accomplished by a “set-up” against the movement of the pointer caused by the thermocouple, so as to prevent any motion over the scale until an assigned temperature is reached. For example, a junction developing 12 millivolts at 1000° C. may be coupled to an indicator in which the full-scale deflection of the pointer is produced by 6 millivolts. If an E.M.F. of 6 millivolts be opposed to the junction, no deflection will occur until the temperature at which the couple develops 6 millivolts is reached—when the opposing E.M.F. will be overcome. This temperature may be 500° C., so that the whole scale may be divided up between 500° and 1000°. The length of the indicator scale is thus effectively doubled; and by using different values for the set-up, it is evident that any desired range may be obtained within the limits of sensitivity of the indicator. The method of procuring the opposing E.M.F. varies with different makers. The Cambridge and Paul Instrument Company employ a dry cell and a series resistance, connected so as to oppose the thermocouple; and by adjusting the resistance any desired set-up may be obtained, the value of which, in degrees, may be read off by connecting the cell and resistance to the indicator, the couple having been switched out of the circuit. Thus, to adjust for a range of 500°-1000° on an indicator giving full-scale deflection for 500°, the resistance is regulated so that the cell alone causes the pointer to move to the end of the scale. The method adopted by Paul consists of suitable resistances inserted in a Wheatstone bridge, which may be thrown off the balance, and thus cause an opposing E.M.F. of the correct amount at the terminals of the indicator.

A mechanical set-up has been introduced by the Cambridge and Paul Instrument Company, the indicator in this case having a suspended coil. By turning a milled-head a twist may be given to the suspending strip, and by the turning of a second head the pointer may be brought back to zero, retaining the initial twist, which is opposed to that produced by the current due to the couple. Thus, if the imposed twist were such as to move the pointer to the 400° mark on the scale, the temperature indicated by the junction would be the observed reading plus 400. By this method it is possible to obtain any desired range within the limits of the indicator. The danger of producing errors due to “creeping” is said to be negligible.

Fig. 19.—Circuit of Northrup’s “Pyrovolter.”

Potentiometer Indicators.—The advantage of measuring E.M.F. by the potentiometer method is that the result is independent of the resistance of the circuit under test, whereas an indicator is affected by changes in the resistance of the circuit in which it is inserted. When long leads are used to connect a couple to its indicator, notable errors may be caused by the varying resistance of the leads, due to changing temperature; and, in addition, the resistance of the couple-wires varies according to temperature and depth of insertion in the furnace. Attempts have therefore been made to produce indicators based on the potentiometer principle, suitable for workshop use, and one form, known as Northrup’s “Pyrovolter,” is arranged as shown in fig. 19, A. A cell D sends a current through a rheostat R, a copper coil C, and a manganin coil S. The copper coil has the same resistance as the copper winding of the indicator G. The couple is connected, with G in circuit, across the manganin coil S, the resistance of this material being unaffected by temperature. By adjusting R until no deflection is shown on G, the drop of volts across S is made equal to the E.M.F. of the couple. To measure this drop, a key is pressed, altering the circuit as shown in B, the indicator being now in series with S and the couple detached. The value of the current passing through S is unchanged, as the indicator coil has the same resistance as the copper coil C, which it now replaces. The deflection on G indicates the value of this current, and, as the drop of volts across S is proportional to the current, G may be marked off to read E.M.F. and the corresponding temperature of the junction. The advantages claimed are that the indicator may be used with any type of junction, and is unaffected by temperature changes in the circuit. A similar instrument is made by the Brown Company of Philadelphia. Up to the present potentiometer indicators have not been adopted to any extent in Britain, and the adjustments necessary to obtain a reading must be accounted a distinct drawback from a workshop standpoint.

Recorders for Thermo-electric Pyrometers.—It is frequently of importance to know not only the existing temperature of a furnace, but also the fluctuations to which it is subject. Continuous observation of a pyrometer would involve too much labour, and it is therefore evident that an automatic recorder would possess many advantages in such cases. A continuous record shows whether the attendant has maintained the temperature between the prescribed limits, and furnishes a permanent history of a given operation, which often serves as a guide to future procedure.

The first successful recorder, suggested by Sir W. Roberts-Austen and designed by Gen. Holden, F.R.S., was used in conjunction with a mirror galvanometer. In its original form, the spot of light from the mirror was made to fall on a sensitized plate, to which a gradual vertical motion was conveyed by connecting the dark slide to a water-float by means of a chain and pulley. The float was placed in a tank of water, which was gradually emptied through a tap, causing the float to sink and the plate to rise. If the deflection of the spot of light remained steady, a vertical straight line was traced on the plate, fluctuations producing a sinuous line. Trials at known temperatures enabled a standard plate to be obtained, divided into degrees, which could be superposed on a trial plate, and the temperatures thus determined. Much valuable work was accomplished with this recorder by Roberts-Austen for the Alloys Research Committee of the Institution of Mechanical Engineers.

Fig. 20.—Roberts-Austen Recorder.

In its modern form (fig. 20) the photographic plate is replaced by a sheet of sensitized paper wound round a drum which rotates at a known rate—say, once in 12 hours—by means of internal clockwork, shown to the left of the figure. The galvanometer is placed at the opposite end, and the mirror is illuminated by means of an electric lamp placed externally, the rays from which are reflected from a prism in the interior on to the mirror. The ray of light leaving the mirror is broken into two portions, one of which passes through a narrow slit on to the sensitized paper, whilst the other portion is reflected on to a ground-glass scale on the lid, divided so as to read temperatures. In this manner the arrangement serves not only as a recorder, but also indicates the existing temperature without necessitating the examination of the sensitized paper. The whole arrangement is made impervious to light, so that it may be used in daylight. A dark room is necessary for fixing the records. When desired, records of two or more pyrometers may be taken on the same sheet, a clockwork device being used to switch each instrument in turn on to the galvanometer for a given period, an external dial indicating which pyrometer is for the time being in circuit.

Whilst it is a drawback to the use of this recorder that the record is not visible, the use of a mirror galvanometer confers a high degree of sensitiveness to the instrument, not possessed by the recorders to be described subsequently.

Fig. 21.—Principle of Thread Recorder.

The Thread Recorder.—In this instrument an intermittent record is secured in ink, possessing the advantages of visibility during the period over which readings are taken, and of permanence without subsequent treatment of the chart. The principle is shown in fig. 21, where A is a boom terminating in a V-shaped piece of ivory, and attached to the galvanometer suspension B. By means of a cam E, rotated by clockwork, a bar D is made to descend at stated intervals, pressing the end of A on to an inked thread G, and causing the thread to touch a paper wound round the drum C. This drum rotates on its axis once in 25 hours by the action of internal clockwork. The continued rotation of the cam E alternately raises and depresses the boom A, leaving it free for a sufficient time to enable it to attain the position it would occupy if the mechanism were absent. The thread G is passed over pulleys, and is wound round through an ink-well, so that the portion opposite A is always moist. With the bar D descending every two minutes, the successive dots form a nearly continuous line. The paper on C is divided horizontally into temperatures, and vertically into time units, so that the temperature existing at any given time may readily be ascertained. The front of the bar D, or a separate strip parallel to it, is divided so as to enable temperatures to be read without reference to the chart. The actual instrument is shown in fig. 22. When several simultaneous records are required, the drum C is extended, and other galvanometers introduced, to which the separate pyrometers are connected. Several records can be taken on one chart by introducing a clockwork mechanism to couple each pyrometer in turn to the one galvanometer.

Fig. 22.—Thread Recorder.


Fig. 23.—Siemens’ Recorder.

The Siemens Recorder.—In this instrument (fig. 23) the boom from the galvanometer terminates in a knife-edge, and moves over a thin horizontal rail, the top of which is rounded. Between the rail and the boom are placed an inking ribbon and a paper chart, which is moved forward by clockwork. A chopper-bar, also actuated by clockwork, descends at about half-minute intervals, and depresses the end of the galvanometer boom, thus producing a small dot on the chart. The paper is 12 cms. wide and 40 yards long; it is divided into time and temperature units, and moves forward at the rate of 2 cms. per hour. Levelling screws are fixed to the base of the recorder.

Fig. 24.—Foster’s Recorder.

Foster’s Recorder.—Foster’s recorder (fig. 24) is designed for use with base-metal couples of the nickel-chromium type, known as Hoskin’s alloys, which yield an E.M.F. about five times as large as a platinum-rhodioplatinum couple. The force available in this case enables the coil of the galvanometer to be pivoted in a horizontal position, the pointer being vertical, and yet to be sufficiently sensitive. The chart is mounted on a vertical plate which rotates on its axis, the time ordinates taking the form of concentric circles, which are cut at an angle by the temperature ordinates. At the terminus of the pointer is placed a small capillary tube, fitted with an inked wick, which, when pressed upon the chart, makes a mark. The presser-bar is curved to the same radius as the pointer, and carries a pad wetted with ink, so that at each depression the supply of ink to the wick is replenished by an amount equal to that imparted to the chart. This recorder is sometimes fitted with special contacts, so that when the correct temperature exists an electric lamp with a white bulb remains lighted; whereas when too low or too high a green or red lamp is lit up, and an alarm thus given. Such an addition involves the use of a relay circuit, but is advisable in cases where expensive articles might suffer if overheated. It can be modified to permit of several simultaneous records being taken, and possesses the advantage that the whole chart is visible at any time. On the other hand, the circular coordinates may be accounted a drawback by some, as not being quite so familiar to read as charts in which the lines are straight. Robust construction is a feature of this recorder.